Olympus exera ii cv-180 manual


















Refer to the respective instruction manual for each piece of equipment. If the power still fails to come ON, contact Olympus. Turn ON the light source and confirm that examination light is emitted from the distal end of the endoscope see Figure 3. For operation of the light source, refer to its instruction manual.

Confirm that this instrument is connected to the light source using the light source cable or light control cable see Section 8.

Please refer to the instruction manual of the endoscope used. Be sure to perform white balance adjustment before inspecting the color on the monitor display. See Section 4. Do not use this instrument when the live image cannot be observed. Otherwise, patient injury may occur. According to Section 5. This chapter explains the work flow of endoscopic observation using the video system center.

For information on how to use the functions that are not explained in this chapter, refer to the reference pages. The operator of the video system center must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in clinical endoscopic techniques.

This manual, therefore, does not explain or discuss clinical endoscopic procedures. It only describes basic operation and precautions related to the operation of the video system center. Then turn it back ON again. For ancillary equipment used in conjunction with the video system center, also turn the power OFF and then ON again as directed in their respective instruction manuals.

If this fails to correct the problem, immediately stop using the equipment and turn the video system center and light source OFF. Then, gently withdraw the endoscope from the patient as described in the endoscope's instruction manual.

If the problems cannot be resolved by the remedial action described in Chapter 10, do not use the equipment and contact Olympus. Electromagnetic radiation can interfere with the monitor display. Non-Olympus equipment can cause interference on the monitor display or a loss of the endoscopic image. If high-frequency electrosurgical equipment is used without such confirmation, patient injury may result. If it is not turned ON, the auto brightness control function is not activated and the light intensity is set to maximum.

In this case, the endoscope distal end would become hot and could cause burns to the operator and physician if a light source model other than CLV is used. Medical agents might enter the video system center through the ventilation grills and may cause equipment damage. Welcome to ManualMachine. We have sent a verification link to to complete your registration. Log In Sign Up. Forgot password?

Enter your email address and check your inbox. Please check your email for further instructions. Enter a new password. Olympus CV User manual Download for 1. Contents PinP picture in picture function Labels and Symbols Labels and Symbols Safety-related labels and symbols are attached on the locations shown below.

Caution that only the exclusive cable can be connected. Potential equalization terminal Electric rating The product name, rated voltage and frequency are shown. Important Information — Please Read Before Use Automatic brightness control: The automatic brightness control automatically adjusts the intensity of the light emitted from the light source so that the endoscopic image will be maintained at constant brightness even if the distance between the distal end of the endoscope's insertion tube and the subject changes.

Color adjustment: Color adjustment adjusts the color balance on the video monitor. Iris: The iris function is used to electrically measure the brightness of an endoscopic image to obtain a control signal for the purpose of automatic light adjustment.

Freeze: The freeze function creates a stationary view of the moving image. Release: The release function is used to capture and record an endoscopic image. Edge enhancement: Edge enhancement is an image processing technique that electronically sharpens the edges of an image. Structure enhancement: Structure enhancement is an image processing technique that electronically emphasizes the detailed patterns and edges of an image to increase sharpness. PinP Picture in picture : PinP function displays both the image of the endoscopic live image and the image of an external device on the monitor simultaneously.

PC card: A digital medium for storage of images, etc. Wash out: Wash out is the inability to see details in the endoscopic image due to excessive brightness. Observe the following warnings on the illumination. Locking lever 3. Video connector socket 2. Power indicator 1. Power switch RESET button Image enhancement mode indicators Iris mode indicators Iris mode IRIS button. NBI indicator 9. HDTV indicator 8. Picture in picture PinP button 5. Image source buttons STOP button PC card status indicator Eject button PC card slot PinP composite terminal.

Chapter 2 Nomenclature and Functions 1. Power indicator Lights up when the video system center is ON. Video connector socket The video plug of the videoscope cable, videoscope or camera head are connected to this socket.

Locking lever Press down to disconnect the video plug of the videoscope cable, videoscope or camera head. Image source buttons Press these buttons to select the image sources to be displayed on the monitor.

Picture in picture PinP button Press to display an image of the connected ancillary equipment and the endoscopic live image together on the monitor.

Chapter 2 Nomenclature and Functions Image enhancement mode indicators One of these indicators light up and indicates the image enhancement mode being selected.

RESET button Press and hold the switch to return the settings changed during operation to the default settings. Chapter 2 Nomenclature and Functions 2. PC remote terminal 7. PC IN terminal 6. PC OUT terminal 5. Printer remote terminal 4. Printer OUT terminal 3. Printer IN terminal 2. Keyboard terminal Potential equalization terminal Monitor OUT terminal Composite OUT terminal Monitor remote terminal PC OUT2 terminal.

Remote terminal Option terminal Foot switch terminal VCR remote terminal Fuse box AC power inlet Light control terminal Light source terminal Digital OUT terminal. Keyboard terminal Connect the keyboard. Printer IN terminal Connect the video printer.

Printer OUT terminal Connect the video printer. Printer remote terminal Connect the video printer. Establishes communication with the video printer. PC IN terminal Connect the image filing system. PC remote terminal Connect the image filing system. Option terminal Reserved for future system expansion. Foot switch terminal Connect the foot switch. Fuse box Stores the fuses that protect the instrument from electrical surges.

AC power inlet Connect the provided power cord to supply the AC power via this inlet. Light control terminal Connect a light source that supports the analog interface. Light source terminal Connect a light source CLV that supports the digital interface. Domepoint Click key PRINT key COLOR key F4 key 8. F8 key 3. F3 key 7. F7 key 2. F2 key 6. F6 key 1. F1 key 5. F5 key. F12 key Pause key F11 key Scroll lock key F10 key Print screen key 9.

F9 key. Chapter 2 Nomenclature and Functions 8. F8 key Press to change the image area size on the monitor.

F9 key Press to switch the image enhancement mode. F10 key Press to stop the VCR. F11 key Press to pause the VCR. F12 key Press to start VCR recording. Print screen key Press to fast-rewind the VCR. Scroll lock key Press to playback the VCR. Pause key Press to fast-forward the VCR. Arrow keys Moves the cursor. Video plug Connect to the video connector socket of the video system center.

Scope side connector Connect to the scope connector of the endoscope. Patient data 2. System clock 3. Image recording device display 4. Image information 5. Flushing pump 6. PC card capacity 7. Index image 8. Attending physician 9. Orientation Car diac end of t he stomach V Scope nickname 1.

Patient data Patient data such as name sex, etc. System clock. Image recording device display The status of the image recording devices that record and print the image are displayed only when the recording devices are activated. Chapter 2 Nomenclature and Functions 4. Image information Displays image information on the monitor. If the problem cannot be resolved using the information in Chapter 10, contact Olympus.

This instrument is to be repaired by Olympus technicians only. Otherwise, explosion or fire may result because this video system center is not explosion-proof. This may impair the performance of the product. Page 17 OFF before connecting or disconnecting the endoscope. Otherwise, the video system center may break down and gets damaged from overheating.

Page 18 Classification types are clearly specified in the instruments' instruction manuals. Page 21 These patient data can be copied to the other CV Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. STOP button Image enhancement PC card status mode ENH. Iris mode indicators Eject button PC card slot Iris mode IRIS button This indicator works only when the light source CLV is used.

Iris mode indicators Indicates the iris mode being selected. Page 26 The indicator goes off when the image enhancement is not used. RESET button Press and hold the switch to return the settings changed during operation to the default settings.

Page Rear Panel Monitor OUT terminal Light source terminal Composite OUT terminal Monitor remote terminal PC OUT2 terminal Page 28 Foot switch terminal Connect the foot switch. Outputs the analog video signal and the remote signals to the VCR. Fuse box Stores the fuses that protect the instrument from electrical surges. Page 29 Connect a monitor compatible with the serial digital interface SDI. Outputs the SDI signal. Digital OUT terminal Connect an Olympus-recommended digital video recorder to output and input the digital video signal to the digital video recorder, using IEEE cable.

Page Keyboard F11 key 7. F7 key Scroll lock key 2. F2 key F10 key 6. F6 key Print screen key 1. F1 key 9. F9 key 5. F5 key Page 33 Press this key to enter an item after selecting the item using the domepoint. Domepoint Moves the arrow pointer. Selects an item in the menu or puts a marking in the endoscopic image.

Page 34 Fixes entry and goes to the next text box or screen. Video plug Scope side connector Connect to the video Connect to the scope connector socket of the connector of the endoscope.

Page Monitor The status of the image recording devices that record and print the image are displayed only when the recording devices are activated. Page 38 Zoom ratio page 91 5. Page 39 Chapter 2 Nomenclature and Functions Scope nickname The scope nickname is displayed when an endoscope with a scope nickname function is connected. Page 40 Scroll bar Shows all setting values not being displayed in the list box.

Arrow pointer Moves the cursor, selects the function buttons, displays the pull down menus. Page Pointer Always displayed on image screen unless full image screen selected. Page Chapter 3 Inspection Prepare the video system center and other ancillary equipment before each particular case. Refer to the respective instruction manual for each piece of equipment.

If the power still fails to come ON, contact Olympus. For operation of the light source, refer to its instruction manual. The indicator above the switch goes off. Then, gently withdraw the endoscope from the patient as described in the endoscope's instruction manual. Page 51 High-frequency electrosurgical equipment can cause slight interference on the monitor display. Page Operation Flow 6. Enter the patient data.

Page 53 8. Terminate the examination. Disconnect the endoscope from the video system center and the light source. Attach the light guide cable, video adapter and camera head to the rigidscope, referring to the instruction manuals for the light guide cable, video adapter and camera head. Rigidscope The numbers show the order of connection. Figure 4. Name: Sex: Age: D. The endoscopic image appears on the monitor and the user preset data is loaded to the video system center.

Page White Balance Adjustment Otherwise, cross- contamination may result. Setting an inappropriate color tone or enhancement condition may result in overlooking or wrong diagnosis.

Table 4. If it is OFF, go back to step 1. When performing white balance adjustment keep an appropriate distance between distal end and white cap. Otherwise, the white balance might not be adjusted properly. Page 63 Hold the endoscope as shown in step 1. Name: Sex: Age: Cursor D. Page 66 When modifying data, press the arrow key to move the cursor to the input position and edit the data. Video plug Scope side connector Scope cable holder Videoscope cable A When the scope side connector B When both connectors are is not connected.

Page 69 Disconnect the video plug of the videoscope cable from the instrument, holding the instrument with a hand so that it will not move and push the locking lever down see Figure 4.

Video connector socket Locking lever Video plug Figure 4. Figure 5. Page 74 Before activating the PinP function make sure that an endoscope is connected to this instrument.

If no endoscope is connected it might not be possible to use the PinP function and image recording of an external device may not work. Page Image Enhancement Mode Enh. The indicator above the button goes OFF.

Page Brightness Adjustment Exposure The light intensity reaches its greatest value, and excessive heat of the endoscope's distal end may occur. Table 5. Page 80 Disconnecting them can increase the light intensity to the maximum and may cause burns or eye injury.

Refer to the instruction manual of the endoscope for details of the electric shutter function. Page 82 Set the brightness as required, using the brightness adjustment buttons of the light source.

The brightness level is displayed on the light source. Press the eject button to eject the PC card after a few seconds, then insert the PC card again to let the video system center recognize the PC card. Do not use a personal computer, etc. PC card status indicator PC card adapter Figure 5. Page 86 PC card slot.

Page 87 Chapter 5 Functions Press the eject button again. The PC card adapter comes out slightly see Figure 5. Eject button PC card adapter Figure 5. Page Reset Button Table 5. All front panel indicators should blink for about 1 second, then the settings are reset. Click the text box to place the cursor. Page 91 Z: x1. Shift Figure 5. For details, see Section 5. For details on the user preset menu, see Section 9.

The color bar appears on the monitor see Figure 5. Page Freeze "Freeze" A frozen image in the frame freeze mode is more easily blurred than in the field freeze mode. The arrow pointer appears in the center of the endoscopic image see Figure 5.

Domepoint Arrow keys Shift Figure 5. Page Ending Examination "Exam End" e. The PC card that is formatted by a personal computer may cause malfunction in recording or playback.

Page 30 images, a message appears on the screen see Figure 5. Page Pc Card Menu See page Balloon 3 is shipped sterile in sets of 20 pieces, enclosed in a resealable package. The correct model to be used with this endoscope is listed in Table 1. Grip section. Biopsy valve MAJ Ultrasonic cable connector.

Connects the ultrasonic cable of the ultrasound diagnostic equipment to the endoscope. Connects the endoscope with the Olympus electrosurgical unit via the S-cord. The S-cord conducts leakage current from the endoscope to the electrosurgical unit. To connect the S-cord, refer to the instruction manual for the electrosurgical unit. Connect the fitting part of the chain for water-resistant cap to this mount, as required see Section 2.

Connects the endoscope to the output socket of the light source and transmits light from the light source to the endoscope. Water supply connector and air supply connector. Connects the endoscope to the water container via the water container tube to supply water to the distal end of the endoscope.

Videoscope cable connector. Connects the endoscope to the video system center via the videoscope cable. The endoscope contains a memory chip that stores information about the endoscope and communicates this information to the video system center CV, CV For more details, refer to the instruction manual for the CV, CV Moving the lock in the opposite direction locks the bending section at any desired position.

Suction valve MAJ The suction valve is depressed to the first stage to activate suction. The valve is also used to remove any fluid or debris adhering to the objective lens. The suction valve is depressed completely to activate suction of sterile water from the balloon through the balloon channel. It also can be used to feed air to remove any fluid or debris adhering to the objective lens.

The valve is depressed completely to fill the balloon with sterile water through the balloon channel. Insertion section limit mark. The elevator moves EndoTherapy accessories when the elevator control lever is operated. The functions of remote switches 1 to 4 can be selected on the video system center.

Refer to the instruction manual for the video system center when setting these functions. This plug is used for connection of the washing tube to clean and disinfect the elevator channel. This code is used to quickly determine the compatibility of EndoTherapy accessories. The endoscope can be used with EndoTherapy accessories that have the same color code. Turning the lock in the opposite direction locks the bending section at any desired position. However, when connecting to an instrument that complies with the EMC standard for medical electrical equipment, edition 1 IEC , the whole system complies with edition 1.

The last digit of the year of manufacture is given in the second digit of the serial number. Doing so may result in the fitting part of the chain detaching from the S-cord connector mount, causing the endoscope to fall. Connecting the fitting part to the suction connector may impair the connection of the suction tube to the suction connector. It may also cause the suction tube to become disconnected from the endoscope and allow patient debris to spray. The water-resistant cap with the chain can only be ultrasonically cleaned if connected to endoscopes.

Otherwise, equipment damage may result. Confirm that the chain for water-resistant cap is free from cracks, flaws, wear, deformation, or other damages see Figure 2.

Align the notch on the connecting plate with the pin on the venting connector of the water-resistant cap MH, see Figure 2. Place the connecting plate over the venting connector see Figure 2. Confirm that the connecting plate is securely attached to the foot of the venting connector and can be smoothly rotated.

The instructions on the remaining pages of this manual are given under the assumption that the chain for water-resistant cap is detached from the endoscope. Before each procedure, prepare and inspect this instrument as instructed below.

Inspect other equipment to be used with this instrument as described in their respective instruction manuals. If this instrument malfunctions, do not use it. Return it to Olympus for repair as described in Section Do not pull the universal cord or the ultrasonic cable with excessive force when the endoscope is connected to the other equipment.

Doing so could cause equipment damage. Prepare the equipment shown in Figure 3. Refer to the respective instruction manuals for each piece of equipment. Then remove the water-resistant cap from the endoscope connector. Inspect the control section and the endoscope connector for excessive scratching, deformation, loose parts, or other irregularities. Inspect the boot and the insertion section near the boot for bends, twists, or other irregularities.

Inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities. Gently holding the insertion section with one hand, carefully run your other hand back and forth over the entire length of the insertion section see Figure 3.

Confirm that no objects or metallic wire protrude from the insertion section. Also, confirm that the insertion tube is not abnormally rigid. Using both hands, bend the insertion tube of the endoscope into a semicircle. Then, moving your hands as shown by the arrows in Figure 3. Gently hold the midpoint of the bending section and a point 20 cm from the distal end.

Push and pull gently to confirm that the junction between the bending section and the insertion tube is not loose. In this case, do not use the endoscope because it may be impossible to straighten the bending section during an examination. Confirm that the bending section angulates smoothly and correctly, that maximum angulation can be achieved, and that the bending section returns to its neutral position.

Move the elevator control lever slowly all the way in the opposite direction of. Visually confirm that the portion of the elevator wire extending from the distal end of the insertion section is not broken or bent see Figure 3. While observing the forceps elevator at the distal end of the insertion. Confirm that the lever can be operated smoothly and that the forceps elevator is raised smoothly. Also confirm that the forceps elevator remains stationary when pushed from behind while holding the elevator control lever stationary see Figure 3.

Confirm that the lever can be operated smoothly and that the forceps elevator is lowered smoothly see Figure 3. Confirm that the holes of the valves are not blocked see Figures 3.

Confirm that the valves are not deformed or cracked see Figures 3. The biopsy valve is a consumable that should be inspected as follows before each use. Replace it with a new one if any irregularity is observed during the inspection. Confirm that the slit and hole on the biopsy valve have no splits, cracks, deformations, discoloration, or other damage see Figure 3.

Do not use a mouthpiece that is damaged, deformed, or reveals other irregularities. Confirm that the mouthpiece is free from cracks, deformations, or discoloration see Figure 3. Using your fingers, check all surfaces of the mouthpiece for scratches, cracks, or other irregularities see Figure 3. Attach the suction valve MAJ to the suction cylinder of the endoscope see Figure 3.

Confirm that valve is fitted properly without any bulging of the skirt. The suction valve will make a whistling noise when it is dry; this does not indicate a malfunction. Spilled water could splash on the equipment, and may cause equipment malfunction.

Prepare and inspect the light source, video system center, ultrasound diagnostic equipment, monitor, water container, suction pump, and EndoTherapy accessories as described in their respective instruction manuals.

Confirm that there are no scratches, cracks, excessive wear, or deformation of the ultrasonic cable. Insert the ultrasonic connector properly into the transducer port of the Olympus universal endoscopic ultrasound center. Insert the ultrasound connector properly into the transducer port of the diagnostic ultrasound system.

If the Ultrasound connector is connected to one of the probe connectors on the upper side PROBE 1 or 2 , the operation panel of the ProSound F75 may hit the Ultrasound connector, which may result in the equipment damage.



0コメント

  • 1000 / 1000